The MRI pre- and post-processing steps that every reconstruction needs and none of them owns: the centred Fourier convention, the receive-array corrections that happen before a solve, the EPI readout corrections, and the shading and apodization that happen after.
The five families, each above the step that corrects it: the centred transform
and its round trip, a receive array's noise covariance before and after
prewhitening, the Nyquist ghost an odd/even phase leaves, Gibbs ringing off a
truncated k-space, and the shading a surface array leaves. Every panel comes
from the same calls the numbered examples make —
python examples/figures/make_showcase.py.
Everything is array in, array out: Torch tensors keep their device and NumPy arrays stay NumPy, so the same step runs on a scanner host and on a workstation without either having to convert to reach it. Torch is an extra, not a dependency.
The package is deliberately small, and it is the base of a family of MRI packages rather than a reconstruction of its own — no solvers, no physics operators, no sensitivity estimation.
pip install mrutils # numpy + scipy
pip install mrutils[torch] # keep tensors on their device
pip install mrutils[bias] # N4 bias-field correction
pip install mrutils[all]The Fourier convention is ifftshift -> fft(norm="ortho") -> fftshift, stated
once so nothing downstream re-derives the shifts, and cropping and padding are
about the same centre so neither moves the object.
import mrutils as mru
# the centred transform, over the last two axes by default
kspace = mru.fftc(image)
plane = mru.ifftc(volume, axes=-1)
# crop or pad about the centre, and drop readout oversampling
coarse = mru.resize_centered(kspace, (128, 128))
readout = mru.remove_readout_oversampling(digitised, 256)
# whiten against a noise-only scan, so the array's coupling stops being a term
whitened = mru.noise_prewhiten(kspace, noise_scan, coil_axis=0)
# compress the array, keeping the basis for every acquisition that follows
compressed, basis = mru.coil_compress(whitened, 8) # or 0.95 of the energy
compressed = mru.apply_coil_compression(basis, later_acquisition)
# stream a scan too large for the card, off the host, for the same answer
compressed = mru.apply_coil_compression(basis, host_scan, device="cuda")
# the EPI Nyquist ghost, measured from a blip-nulled navigator
phase = mru.estimate_epi_phase([plus, minus, plus])
lines = mru.correct_lines(train, phase) # [(data, is_reversed), ...]
# ramp sampling, as a change of basis rather than an interpolation
operator = mru.epi_ramp_operator(sampled_k, uniform_k, support=matrix)
on_grid = readout @ operator.T
# apodize: radial for isotropic resolution, separable to keep the corners
apodized = mru.apodize(kspace, kind="fermi")
slab = mru.apodize(volume, kind="hann", geometry="separable", axes=(-3, -2, -1))
window = mru.fermi_window((256, 256), radius=0.9, width=0.05)
# N4 on the coil-combined magnitude, after the solve
corrected = mru.bias_field_correct(magnitude) # needs [bias]Each runs its function against something known: a round trip, a covariance, a
phase impressed on purpose, a field with a closed form. The .py beside each
notebook is the source — it runs as a script and lints with the rest of the
package, and scripts/build_examples.sh is what turns it into the notebook.
01-fourier |
the centred convention, and what cropping about the wrong centre does | |
02-coils |
prewhitening a receive array, and compressing it | |
03-epi |
the Nyquist ghost, and ramp sampling | |
04-apodization |
windows against Gibbs ringing, radial against separable | |
05-bias_field |
N4 against the shading a surface array leaves |
- ismrmrd-python-tools — https://github.com/ismrmrd/ismrmrd-python-tools. The same idea: the reconstruction steps that are common to everyone, as plain array functions. This package keeps that scope and drops the ISMRMRD file format, so nothing here needs a data model to be called.
- SimpleITK — https://simpleitk.org/. Its
N4BiasFieldCorrectionImageFilteris whatbias_field_correctcalls. - Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array. Magn Reson Med 1990;16:192-225. The noise covariance and the whitening built from it.
- Buehrer M, Pruessmann KP, Boesiger P, Kozerke S. Array compression for MRI with large coil arrays. Magn Reson Med 2007;57:1131-1139.
- Bruder H, Fischer H, Reinfelder HE, Schmitt F. Image reconstruction for echo planar imaging with nonequidistant k-space sampling. Magn Reson Med 1992;23:311-323. The odd/even phase correction and the ramp-sampling resampling.
- Bernstein MA, Fain SB, Riederer SJ. Effect of windowing and zero-filled reconstruction of MRI data on spatial resolution and acquisition strategy. J Magn Reson Imaging 2001;14:270-280. The window kernels, the half-height-at- Nyquist normalization, and the radial-versus-separable geometry — whose 52.4% (2D) and 50.7% (3D) corner ratios the examples reproduce.
- Tustison NJ, Avants BB, Cook PA, Zheng Y, Egan A, Yushkevich PA, Gee JC. N4ITK: improved N3 bias correction. IEEE Trans Med Imaging 2010;29:1310-1320.
pip install -e .[dev]
bash scripts/format_and_lint.sh
pytest -q
bash scripts/build_examples.sh # rebuild the notebooks and their figuresThe docstring examples run as part of the suite — they are the documentation, and an example that has drifted is a broken one. See CONTRIBUTING.md.